Glass electrode calibration for use in the voltammetric determination of stability constants under extreme acidic conditions
dc.contributor.author | Billing, Caren | |
dc.contributor.author | Cukrowski, Ignacy | |
dc.date.accessioned | 2010-04-06T07:02:17Z | |
dc.date.available | 2010-04-06T07:02:17Z | |
dc.date.issued | 2009-09 | |
dc.description.abstract | A glass electrode (GE) can be successfully employed to measure pH in the study of metal-ligand equilibria by voltammetry at extremely low pH (between 0 and 2); two consecutive strong acid–strong base titrations involving different base concentrations (recommended to avoid corrosion of the GE in very basic solutions) are best suited to establish the response parameters of a GE. A novel approach of using a combined linear and binomial GE calibration was developed; this procedure allows measurements between pH 0 and 2 with uncertainty better than ±0.01 pH unit. From an extensive error analysis, it has been established that the uncertainties of about ±0.5 mV in the response slope and ±1.3 mV in E°’ might result in an absolute error in pH of about 0.02 which should not generate errors larger than 0.3%in optimized stability constants (as logKvalues) determined by voltammetry at extremely low pH. A test of GE suitability for the study of metal complexes by voltammetry is also proposed; it should be implemented only for suspect electrodes that show response parameters outside the limits recommended in this work. | en |
dc.identifier.citation | Billing, C & Cukrowski, I 2009, 'Glass electrode calibration for use in the voltammetric determination of stability constants under extreme acidic conditions', South African Journal of Chemistry, vol. 62, pp. 168-178. [http://journals.sabinet.co.za/sajchem/] | en |
dc.identifier.issn | 0379-4350 | |
dc.identifier.uri | http://hdl.handle.net/2263/13774 | |
dc.language.iso | en | en |
dc.publisher | South African Chemical Institute | en |
dc.rights | South African Chemical Institute | en |
dc.subject | Glass electrode calibration | en |
dc.subject | Glass electrode performance | en |
dc.subject | Metal-ligand equilibria studies | en |
dc.subject.lcsh | Electrodes, Glass -- Calibration | en |
dc.subject.lcsh | Voltammetry | en |
dc.subject.lcsh | Potentiometry | en |
dc.title | Glass electrode calibration for use in the voltammetric determination of stability constants under extreme acidic conditions | en |
dc.type | Article | en |