Abstract

A multimodal sensor was developed to record dissolved oxygen, L*a*b* colour, temperature, and pH. This work builds on an existing model that correlates blood oxygen saturation with L*a*b* colour values. An L*a*b* colour sensor was constructed from an RGB sensor and validated against a commercial colourimeter. Sensor performance was confirmed using reference colours. Dissolved oxygen was measured with a screen-printed electrode and an analogue-to-digital converter. The results highlight potential for future optical determination of oxygen saturation, combined with electrochemical measurement of oxygen partial pressure, and compensation for pH and temperature.

Description

DATA AVAILABILITY STATEMENT : Raw data for L*a*b* and concentration are included in this article. For the pH and temperature, only summarized data (mean ± standard deviation) are reported. Additional raw data are available from the corresponding author upon reasonable request. NOTE : Presented at the Micro Manufacturing Convergence Conference, Stellenbosch, South Africa, 7–9 July 2024.

Keywords

Blood oxygen saturation, Screen printed electrodes, L*a*b*, Dissolved oxygen, Temperature sensor

Sustainable Development Goals

SDG-03: Good health and well-being
SDG-09: Industry, innovation and infrastructure

Citation

Mpala, K. & Joubert, T.-H. Veterinary Blood Oxygen Detection. Engineering Proceedings 2025, 109, 20: 1-5. https:// doi.org/10.3390/engproc2025109020.