Thermocouples for temperature measurements on skin surfaces and in living tissue
dc.contributor.author | Elder, J.S. | |
dc.contributor.editor | Du Toit, P.J. | |
dc.date.accessioned | 2017-02-23T11:46:26Z | |
dc.date.available | 2017-02-23T11:46:26Z | |
dc.date.created | 2017 | |
dc.date.issued | 1941 | |
dc.description | The articles have been scanned in colour with a HP Scanjet 5590; 300dpi. Adobe Acrobat XI Pro was used to OCR the text and also for the merging and conversion to the final presentation PDF-format. | en_ZA |
dc.description.abstract | Details have been given of the construction of thermocouples suitable for temperature measurements on skin surfaces and in living tissue. The sources of error have been discussed, and tests of the accuracy of the method have been described. It is concluded that, with adequate precautions, the following degree of accuracy was attained : Skin Surface Instrument - Random errors, ±0·14°C.; systematic errors, less than - 0·2°C. per 10°C. temperature difference between skin and surroundings. Hypodermic Instrument - Random errors, ± 0·1°C.; systematic errors, less than 0·1°C. per 10°C. temperature difference between tissue and air. | en_ZA |
dc.identifier.citation | Elder, JS 1941, 'Thermocouples for temperature measurements on skin surfaces and in living tissue', Onderstepoort Journal of Veterinary Science and Animal Industry, vol. 17, nos. 1 & 2, pp. 141-163. | en_ZA |
dc.identifier.issn | 0330-2465 | |
dc.identifier.uri | http://hdl.handle.net/2263/59148 | |
dc.language.iso | en | en_ZA |
dc.publisher | Pretoria : The Government Printer | en_ZA |
dc.rights | © 1941 ARC - Onderstepoort and Faculty of Veterinary Science, University of Pretoria (original). © 2017 University of Pretoria. Dept. of Library Services (digital). | en_ZA |
dc.subject.lcsh | Veterinary medicine -- South Africa | |
dc.title | Thermocouples for temperature measurements on skin surfaces and in living tissue | en_ZA |
dc.type | Article | en_ZA |