Chirp-evoked VEMPs : a test-retest reliability study

Show simple item record

dc.contributor.author Reddy, Tarryn Marisca
dc.contributor.author Heinze, Barbara
dc.contributor.author Biagio-de Jager, Leigh
dc.contributor.author Maes, Leen
dc.date.accessioned 2024-02-09T08:51:03Z
dc.date.issued 2024-01
dc.description.abstract OBJECTIVES : To determine the test-retest reliability of cervical and ocular vestibular-evoked myogenic potentials (c&oVEMP) evoked by 500 Hz narrowband (NB) CE-Chirp and broadband (BB) CE-Chirp stimuli. DESIGN : Twenty healthy participants (10 female) were tested twice on the same day to determine the within-session reliability and 1 week later to determine the between-session reliability. The latency, amplitude, and asymmetry ratio of c&oVEMPs elicited by 95 dB nHL air-conducted (AC) 500 Hz NB CE-Chirp and BB CE-Chirp were recorded bilaterally. RESULTS : A moderate to good between-session reliability with intraclass correlation coefficient (ICC) values ranging from 0.52 to 0.82 was observed for cVEMP latency, amplitude, and asymmetry ratio evoked by 500 Hz NB CE-Chirp, as well as for the BB CE-Chirp cVEMP amplitude (ICC of 0.70 and 0.84). In contrast, an overall poor reliability ICC values between 0.30 and 0.42 for cVEMP latency and asymmetry ratio were observed for BB CE-Chirp. For the oVEMP, overall poor between-session reliability for all response parameters evoked by the 500 Hz NB CE-Chirp and the BB CE-Chirp was observed. CONCLUSIONS : The 500 Hz NB CE-Chirp was more reliable than the BB CE-Chirp in terms of cVEMP latency, amplitude, and asymmetry ratio. Further investigation using the standard electrode montage is necessary to assess the test-retest reliability of the chirp-evoked oVEMP. en_US
dc.description.department Speech-Language Pathology and Audiology en_US
dc.description.embargo 2024-08-15
dc.description.librarian hj2024 en_US
dc.description.sdg None en_US
dc.description.uri https://journals.lww.com/ear-hearing/pages/default.aspx en_US
dc.identifier.citation Reddy, T.M., Heinze, B., Biagio-De Jager, L. et al. 2024, 'Chirp-evoked VEMPs : a test-retest reliability study', Ear and Hearing, vol. 45, no. 1, pp. 207-218, doi : 10.1097/AUD.0000000000001416. en_US
dc.identifier.issn 0196-0202 (print)
dc.identifier.issn 1538-4667 (online)
dc.identifier.other 10.1097/AUD.0000000000001416
dc.identifier.uri http://hdl.handle.net/2263/94406
dc.language.iso en en_US
dc.publisher Lippincott Williams and Wilkins en_US
dc.rights © 2023 Wolters Kluwer Health, Inc. All rights reserved. This is a non-final version of an article published in final form in Ear and Hearing, vol. 45, no. 1, pp. 207-218, doi : 10.1097/AUD.0000000000001416. en_US
dc.subject Cervical and ocular vestibular-evoked myogenic potentials (c&oVEMP) en_US
dc.subject Cervical VEMP en_US
dc.subject Ocular VEMP en_US
dc.subject Test-retest reliability en_US
dc.subject CE-Chirp en_US
dc.subject Saccule en_US
dc.subject Utricle en_US
dc.subject Reproducibility en_US
dc.subject Vestibular-evoked myogenic potentials (VEMP) en_US
dc.title Chirp-evoked VEMPs : a test-retest reliability study en_US
dc.type Postprint Article en_US


Files in this item

This item appears in the following Collection(s)

Show simple item record