Abstract:
BACKGROUND : Late latency auditory evoked potentials (LLAEPs) provide objective
evidence of an individual’s central auditory processing abilities. Electrically evoked cortical
auditory evoked potentials (eCAEPs) are a type of LLAEP that provides an objective
measure of aided speech perception and auditory processing abilities in cochlear implant
(CI) recipients.
AIM : To determine the short-term test-retest reliability of eCAEPs in adult CI recipients.
DESIGN : An explorative, within-subject repeated measures research design
was employed.
STUDY SAMPLE : The study sample included 12 post-lingually deafened, unilaterally
implanted adult CI recipients with at least 9 months of CI experience.
METHOD : eCAEPs representing basal, medial and apical cochlear regions were recorded
in the implanted ears of each participant. Measurements were repeated 7 days after the
initial assessment.
RESULTS : No significant differences between either median latencies or amplitudes at
test and retest sessions (p > 0.05) were found when results for apical, medial and
basal electrodes were averaged together. Mean intraclass correlation coefficient (ICC)
scores averaged across basal, medial and apical cochlear stimulus regions indicated that
both consistency and agreement were statistically significant and ranged from moderate
to good (ICC = 0.58–0.86, p < 0.05). ICC confidence intervals did demonstrate
considerable individual variability in both latency and amplitudes.
CONCLUSION : eCAEP latencies and amplitudes demonstrated moderate to good
short-term test-retest reliability. However, confidence intervals indicated individual variability in measurement consistency which is likely linked to attention and listening effort required from the CI recipients.