Implant failure due to titanium hypersensitivity/allergy? - report of a case

Show simple item record

dc.contributor.author Du Preez, L.A.
dc.contributor.author Butow, Kurt-Wilhelm
dc.contributor.author Swart, Theuns J.P.
dc.date.accessioned 2008-02-29T12:27:09Z
dc.date.available 2008-02-29T12:27:09Z
dc.date.issued 2007-02
dc.description.abstract Six titanium implants were placed in the mandible between the left and right mental foramen of a 49-year old female patient. The implants were all constructed of the same grade IV titanium. Three types of implants were used, namely LIBB compression implant, cylindrical implant and Brånemarklike implant. These three types of implants were placed as part of a research project, which received approval from the ethics committee. The patient developed a severe reaction to all the implant units, with both clinical and radiological features of complication. The localised tissue reaction was severe enough to warrant removal of all the implants. The surrounding soft tissue was submitted for histological analysis, which revealed a chronic inflammatory response with concomitant fibrosis around all the implants as well as foreign body giant cell reaction around two implants. Following implant removal the patient recuperated well and the soft and hard tissue healed satisfactorily. Following is a case report of a possible true titanium allergy in a clinical setting. en
dc.format.extent 771598 bytes
dc.format.mimetype application/pdf
dc.identifier.citation Du Preez, LA, Bütow, KW & Swart, TJP 2007, 'Implant failure due to titanium hypersensitivity/allergy? – report of a case', South African Dental Journal, vol. 62, no. 1, pp. 22, 24-25. [www.sadanet.co.za] en
dc.identifier.issn 1029-4864
dc.identifier.uri http://hdl.handle.net/2263/4623
dc.language.iso en en
dc.publisher South African Dental Association en
dc.rights South African Dental Association en
dc.subject Dental titanium implants en
dc.subject Titanium hypersensitivity en
dc.subject Titanium allergy en
dc.subject.lcsh Dental metallurgy
dc.subject.lcsh Stomatitis
dc.subject.lcsh Implants, Artificial -- Complications
dc.title Implant failure due to titanium hypersensitivity/allergy? - report of a case en